Operation
52
9.4 General information
Fig. 9.2:
General information
Display
(example values)
Information
V 4.39R09
Version number of the software
04.02.2015 10:05:09
Current date / time
Methane: 95.01 Vol%
Volume share of methane
mV=24.56
mV signal of the thermal battery
V=1.071
Voltage signal of the density measuring cell
T1=29.04/0.24
Inlet temperature into the thermal body / heating of the
air
p=4.00
Air differential pressure
Fr=51.13
Controlled fan frequency
Stabi=0.011
Standard deviation of the measurement value over 2
minutes
NOTE
STD: During a calibration, 0.015 is typically achieved. The calibration is
concluded after that.