Transport, installation, and acceptance
30
6.3.2 Room ventilation
WARNING
Risk of injury due to temperature of exhaust gas/housing!
The exhaust gas temperature is between 8 - 20 °C above the housing tempera-
ture!
Use personal safety equipment to prevent burns!
NOTE
In the case of unfavorable circulation/installation conditions, appropriate deflect-
ing sheets must be provided that prevent fresh air from flowing directly onto the
combustion calorimeter.
The following conditions must be met:
•
No direct fresh air introduction in the combustion calorimeter.
•
Exhaust gases must be discharged to the environment protected from air
draft using a suitable pipe/tube and with the help of corresponding fresh air
mixing. Connection points must be checked for leak tightness.