Transport, installation, and acceptance
41
Digital control inputs Connector X3
Control inputs
Pin/Connector
X3
Function
Status display diode
1
1
2
Start measurement
Start measurement
D 25
2
3
4
Start calibration
Start calibration
D 24
3
5
6
Function, if applica-
ble
optionally assigned
D 23
4
7
8
Function, if applica-
ble
optionally assigned
D 22
5
9
10
Function, if applica-
ble
optionally assigned
D 21
6
11
12
Function, if applica-
ble
optionally assigned
D 20
7
13
14
Function, if applica-
ble
optionally assigned
D 19
8
15
16
Function, if applica-
ble
optionally assigned
D 18
Serial interface RS232
Pin/Connector
X11
Signal
1
-mA analog input 1
2
+mA analog input 1
3
-mA analog input 2
4
+mA analog input 2
5
PT 100 air conditioner
6
PT 100 air conditioner
7
Not assigned
8
RI
9
RTS
10
CTS
11
DSR
12
DTR
13
TXD
14
RXD
15
DCD
16
RS232 GND
Pin/Connector
X1
Signal
1
DCD
2
RXD
3
TXD
4
DTR
5
RS232 DND
6
DSR
7
RTS
8
CT
9
RI
Summary of Contents for CWD2005 SPC
Page 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Page 10: ...Technical data 10 ...
Page 12: ...EU Declaration of Conformity 12 ...
Page 18: ...Safety information 18 ...
Page 22: ...Protective equipment 22 ...
Page 48: ...Description of the work stations operator control elements 48 ...
Page 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Page 87: ...Operation 87 ...
Page 88: ...Operation 88 ...
Page 90: ...Decommissioning Switching off 90 ...
Page 94: ...Maintenance 94 ...
Page 100: ...Troubleshooting 100 ...
Page 102: ...Service 102 ...
Page 104: ...Disposal 104 ...
Page 106: ...Spare parts 106 ...