9.4
Power Failures
73
9
9.4.1 Power Failure Counter-operation
Table 9-3 shows the counter-operations against power interruption which occurred while the
pump rotor is normally rotating.
Table 9-3
Counter-operations Against Power Supply Failure
(Note 1) The time can be changed to 2 seconds from 1 second via communication operation. Refer to
separate manual for serial communication.
(Note 2) For restarting after the power re-supply (restoration), push the RESET button.
(Note 3) The pin numbers are shown in Fig. 6-4 "Arrangement of Remote-control Connector Pins".
(Note 4) Actions can be changed via communication operations. Refer to Section 6.7 "Remote-control
Connector".
Interruption time
1 second or less (Note 1)
Over 1 second (Note 1)
Interrupt/re-supply
During interruption
After re-supply
During interruption
After re-supply
Pump status
Magnetic levitation
Levitation goes on
Levitation goes on
Levitation goes on
Levitation goes on
Run
Decelerates
Returns to before
power-failure running
condition
Decelerates
Decelerates
(Note 2)
Indicator lamp
ROTATION
Lamp ON goes on
Lamp ON goes on
Lamp ON goes on
Lamp ON goes on
STATUS
Before-power-failure
indication goes on
Before-power-failure
indication goes on
Orange lamp turns on Orange lamp ON
goes on
Remote-control output signals (Note 3)
ROTATION
Pin no. (17)-(19)
(18)-(19)
"CLOSE" goes on
"OPEN" goes on
"CLOSE" goes on
"OPEN" goes on
"CLOSE" goes on
"OPEN" goes on
"CLOSE" goes on
"OPEN" goes on
NORMAL
Pin no. (11)-(13)
(12)-(13)
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Contact open
Contact close
"OPEN" goes on
"CLOSE" goes on
ACCELERATION
Pin no. ( 8 )-(10)
( 9 )-(10)
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Contact open
Contact close
"OPEN" goes on
"CLOSE" goes on
BRAKE
Pin no. (14)-(16)
(15)-(16)
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Contact close
Contact open
"CLOSE" goes on
"OPEN" goes on
REMOTE
Pin no. (20)-(22)
(21)-(22)
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Before-power-failure
condition goes on
Contact open
Contact close
Return to Before
power-failure
Return to Before
power-failure
ALARM (Note 4)
Pin no. (27)-(28)
(26)-(28)
"CLOSE" goes on
"OPEN" goes on
"CLOSE" goes on
"OPEN" goes on
Contact open
Contact close
By resetting
Contact close
Contact open
WARNING (Note 4)
Pin no. (29)-(31)
(30)-(31)
"OPEN" goes on
"CLOSE" goes on
"OPEN" goes on
"CLOSE" goes on
"OPEN" goes on
"CLOSE" goes on
By resetting
"OPEN" goes on
"CLOSE" goes on
Summary of Contents for UTM2300 Series
Page 2: ...This page is intentionally left blank...
Page 10: ...viii Location of Label...
Page 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Page 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Page 30: ...16 3 3 3 Controller This page is intentionally left blank...
Page 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Page 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Page 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Page 73: ...7 7GAS PURGE...
Page 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Page 95: ...10 10WARRANTY CLAUSES...
Page 98: ...84 10 This page is intentionally left blank...
Page 100: ...Index Index 2 This page is intentionally left blank...