SECTION 9
TROUBLESHOOTING
72
9
9.4
Power Failures
When a power interruption occurs, the motor inside the turbo molecular pump immediately
begins regenerative braking. The magnetic bearing will use this generated electricity to keep
functioning and the rotor inside the turbo molecular pump will continue being levitated. The
rotation will slow down due to the regenerative braking and eventually the rotor will be supported
by the touchdown bearing. Table 9-2 shows the number of revolutions and period of time that will
elapse before being supported by the touchdown bearing, when the power fails at the rated
speed.
Table 9-2
Time and Rotational Speed During a Power Interruption Before being Supported
by Touchdown Bearing
(Note 1) The time is typical for regenerative braking from the rated speed.
Actual time will vary depending on vacuum conditions inside the pump and the rotational
speed when the power fails.
Pump model
Rotational speed before support
by touchdown bearing
Period of time before support
by touchdown bearing (Note 1)
UTM2300 Series
7500 rpm
9 minutes or less
Summary of Contents for UTM2300 Series
Page 2: ...This page is intentionally left blank...
Page 10: ...viii Location of Label...
Page 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Page 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Page 30: ...16 3 3 3 Controller This page is intentionally left blank...
Page 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Page 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Page 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Page 73: ...7 7GAS PURGE...
Page 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Page 95: ...10 10WARRANTY CLAUSES...
Page 98: ...84 10 This page is intentionally left blank...
Page 100: ...Index Index 2 This page is intentionally left blank...