Alarm Detection Capabilities
NOTICE
Alarm history data can be read via serial interface, but cannot be read via remote control
connector.
The fault detection functions shown in Table 9-6 "Table of Alarms" and Table 9-7 "Table of
Warnings" are incorporated for protection in the event of a problem with the turbo molecular
pump or power supply unit.
When an error is detected, check the STATUS lamp (Fig. 2-2 (7)) and Refer to Section 9.5.3
"Reset Procedure" for the appropriate remedy.
9.5.1 Movement in Alarm Detection Capabilities (ALARM)
1. STATUS orange lamp (Fig. 2-2 (7)) lights.
2. The remote-control connector "ALARM" signal turn on.
3. The pump start the protective operations shown in Table 9-6 "Table of Alarms".
4. The detection error is recorded in the error log.
9.5.2 Movement in Alarm Detection Capabilities (WARNING)
1. STATUS orange lamp (Fig. 2-2 (7)) flashes.
2. The remote-control connector "WARNING" signal turn on.
3. Pump operation continues.
4. The detection error is recorded in the error log.
9.5.3 Reset Procedure
1. Refer to the Troubleshooting information and eliminate the cause of the problem.
2. Conduct the reset operation.
3. (ALARM) If the problem has been eliminated, the STATUS orange lamp (Fig. 2-2 (7)) goes
out, the "ALARM" remote-control signals (Refer to Table 6-4) turn off after an alarm was
given, the pump rotor decelerates rotational speed.
(WARNING) When the problem is eliminated after a warning occurred, pump operation
continues.
4. If the problem was not completely eliminated, the alarm is not cancelled.
Summary of Contents for UTM2300 Series
Page 2: ...This page is intentionally left blank...
Page 10: ...viii Location of Label...
Page 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Page 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Page 30: ...16 3 3 3 Controller This page is intentionally left blank...
Page 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Page 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Page 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Page 73: ...7 7GAS PURGE...
Page 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Page 95: ...10 10WARRANTY CLAUSES...
Page 98: ...84 10 This page is intentionally left blank...
Page 100: ...Index Index 2 This page is intentionally left blank...