47
13. Special test mode
Model
8527
is able to have the setting of 4 special functions by means of key operation on
the front panel.
(1) Double action start function
Within 0.5 second from the stop signal having been input, the test starts with input of
start signal.
Note
: When the function is set,
READY
lamp blinks in READY status.
(2) GOOD hold function
This is the function to concern the good judgement. The output becomes continuous
until the stop signal is input.
(3) Momentary start function
The test is done only when the start signal is input.
(4) FAIL mode function
This is the function to disable the resetting of NG judgement and PROTECTION
action by the stop signal of remote control, and enables the resetting only by the stop
switch on the tester main unit.
Setting procedure of special test mode
①
In READY status, press
SHIFT
and
STOP
key at a time for 3 seconds or more.
READY
lamp blinks and the test time display is lit up with “
”.
The 4th digit of the current/resistance display blinks.
②
The item to set can be moved with
or
key.
③
Refer to the following table for the items to select.
CURRENT/RESISTANCE
▲
key: Numeral increases.
▼
key: Numeral decreases.
Lamps to synchronously blinks
at the setting
-
-
-
Cancel of setting
READY
lamp
-
-
-
Setting of double action start function
-
-
Cancel of setting
GOOD
of
I-TEST
,
W-TEST
-
-
Setting of GOOD hold function
Note
: In order to re-start, once of stop
signal input is necessary
-
-
Setting of GOOD hold function
Note
: When the start signal is input, the
judgement output is reset and
re-starts.
-
Cancel of setting
I-TEST
,
W-TEST
lamp
-
Setting of momentary start function
Cancel of setting
HIGH
LOW
of
I-TEST
,
W-TEST
Setting of FAIL mode
Finish of setting
Press
ENTER
key, then the tester returns to READY status, memorizing the set having
been made.
When the
EXIT
key (
SHIFT
and
ENTER
at a time) is pressed in the setting mode,
the special test mode is interrupted and the tester becomes READY status.
The special test mode in this case is that before entering the special test mode.
ON/OFF
ENTER
WRITE
READ
LOCK
○
○
BUZZER
W
READY
REMOTE
DISCHARGE
KEY LOCK
EXIT
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
SHIFT
M
Ω
LOW
HIGH
LOW
HIGH
LOW
○
W-TEST
○
○
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
HIGH
mA
○
○
○
○
I
I-TEST
I
I-W
W-I
◎
GOOD
GOOD
Double action
GOOD hold
Momentary
FAIL mode