28
Explanation of command
4.2.38
DATA?
(lump read-out of test result)
Function
Reads out the detail data of test result.
[Command to use after the finish of the test (READY status)]
Judgement result and data are retained until the next start even if the
RESET command is made or
STOP
switch is pressed.
Structure
DATA?
Transmission
DATA?
Response
[Example of responses after the finish of automatic test]
Test mode
Judgement result and action
during the test
Response
WI or IW Passed W and I test.
JUDGE=GOOD, WJUDGE=GOOD, VOLT=1.51kV, CURRENT=1.23mA,
IJUDGE=GOOD, RESISTANCE=123MOHM
W test failed for HIGH
JUDGE=NG, WJUDGE=HIGH, VOLT=1.51kV, CURRENT=32.1mA,
IJUDGE=NULL, RESISTANCE=0.00MOHM
W test failed for LOW
JUDGE=NG, WJUDGE=LOW, VOLT=1.51kV, CURRENT=0.15mA,
IJUDGE=NULL, RESISTANCE=0.00MOHM
I test failed for HIGH
JUDGE=NG, WJUDGE=GOOD, VOLT=1.51kV, CURRENT=1.23mA,
IJUDGE=HIGH, RESISTANCE=1234MOHM
WI
I test failed for LOW
JUDGE=NG, WJUDGE=GOOD, VOLT=1.51kV, CURRENT=1.23mA,
IJUDGE=LOW, RESISTANCE=10.20MOHM
W test failed for HIGH
JUDGE=NG, WJUDGE=HIGH, VOLT=1.51kV, CURRENT=32.1mA,
IJUDGE=GOOD, RESISTANCE=12.34MOHM
W test failed for LOW
JUDGE=NG, WJUDGE=LOW, VOLT=1.51kV, CURRENT=0.6mA,
IJUDGE=GOOD, RESISTANE=1234MOHM
I test failed for HIGH
JUDGE=NG, WJUDGE=NULL, VOLT=0.00kV, CURRENT=0.00mA,
IJUDGE=HIGH, RESISTANCE=1234MOHM
IW
I test failed for LOW
JUDGE=NG, WJUDGE=NULL, VOLT=0.00kV, CURRENT=0.00mA,
IJUDGE=LOW, RESISTANCE=9.99MOHM
At RESET (stop)
Note-1
JUDGE=NULL, WJUDGE=NULL, VOLT=0.00kV, CURRENT=0.00mA,
IJUDGE=NULL, RESISTANCE=0.00MOHM
WI or IW When protective function
activated
Occurred in W of WI mode
Note-2
JUDGE=PROTECT, WJUDGE=HIGH LOW, VOLT=1.50kV,
CURRENT=1.23mA, IJUDGE=NULL, RESISTANCE=0.00MOHM
W test passed
JUDGE=GOOD, WJUDGE=GOOD, VOLT=1.51kV,
CURRENT=1.23mA
W test failed for HIGH
JUDGE=NG, WJUDGE=HIGH, VOLT=1.51kV, CURRENT=32.1mA
W test failed for HIGH
When the leak current
exceeds 199.9mA
JUDGE=NG, WJUDGE=HIGH, VOLT=0.00kV, CURRENT=OVER
W test failed for LOW
JUDGE=NG, WJUDGE=LOW, VOLT=1.51kV, CURRENT=0.15mA
At RESET (stop)
Note-1
JUDGE=NULL, WJUDGE=NULL, VOLT=0.00kV,
CURRENT=0.0mA
W
When protective function
activated
Note-2
JUDGE=PROTECT, WJUDGE=HIGH LOW, VOLT=1.50kV,
CURRENT=1.23mA
I test passed
JUDGE=GOOD, IJUDGE=GOOD, RESISTANCE=1234MOHM
I test passed
When it exceeds 2000M
Ω
JUDGE=GOOD, IJUDGE=GOOD, RESISTANCE=OVER
I test failed for HIGH
JUDGE=NG, IJUDGE=HIGH, RESISTANCE=1234MOHM
I test failed for LOW
JUDGE=NG, IJUDGE=LOW, RESISTANCE=1.2MOHM
At RESET (stop)
Note-1
JUDGE=NULL, IJUDGE=NULL, RESISTANCE=0.00MOHM
I
When protective function
activated
Note-2
JUDGE=PROTECT, IJUDGE=HIGH LOW, RESISTANCE=1234MOHM
Note-1
: Data is 0.
Note-2
: Responses with the data when the protective function is activated.
For the test which could not be performed, the data is 0.