36
Setting of test condition for insulation resistance test
[When setting the test time]
To enter setting mode
①
In READY status, press
or
key, then the test mode lamp blinks.
Test mode lamp moves up and down with
▲
or
▼
key. Make the required test
mode lamp blinking (
W-I
,
I-W
or
I
lamp).
②
Press
or
key and make the test mode lamp lit.
Setting of test time
①
Press
or
key and select the status that the test time display blinks.
②
Next, press
ON/OFF
key and select the status that the display blinks with the
numeral (refer to the above figure).
③
Press
▲
or
▼
key and set the test time.
Pressing of
key (
SHIFT
and
▲
at a time) or
key (
SHIFT
and
▼
at a time) allows the setting of second digit (digit of 1s).
To return to the previous setting
Press
key, then changes to the
setting of mask timer time
.
To the next setting
Press
key, then changes to the
setting of discharging function
.
Finish of setting
Press
ENTER
key, then the tester returns to READY status, memorizing the set having
been made.
The slider moves up or down when the test mode is W-I or I-W.
When the
EXIT
key (
SHIFT
and
ENTER
at a time) is pressed in the setting mode,
the setting mode for the test condition is interrupted and the tester becomes READY
status.
The test conditions in this case are those before entering the setting mode of test condition.
ON/OFF
ENTER
WRITE
READ
LOCK
○
○
BUZZER
W
READY
REMOTE
DISCHARGE
KEY LOCK
EXIT
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
SHIFT
GOOD
HIGH
LOW
GOOD
HIGH
LOW
○
W-TEST
○
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
HIGH
mA
○
○
○
I
I-TEST
I
I-W
LOW
M
Ω
●
W-I
●
●
VOLTAGE SLIDER