56
Status output
16.3
●
Setting of condition for status output
Setting procedure of condition for status output
①
Press
ON/OFF
key and
▲
key at a time for 3 seconds or more.
W AC
,
I DC
lamps blink and the test time display is lit up with “
”.
The highest digit of the voltage display blinks.
②
The item to set can be moved with
or
key.
③
Refer to the following table for the items to select.
“ ” or “ ” blinks at selection of each item.
: To select
: Not to select
key:
Move to right, but when pressed at the item
POWER ON, moves to the TEST/H.V. OUT.
key:
Move to left, but when pressed at the item
TEST/H.V. OUT, moves to POWER ON.
Selection item
of status output
Lamps to blink at the setting
TEST/H.V.OUT
DANGER
,
W AC
,
I DC
lamp
TEST
I-TEST
,
W-TEST
lamp
GOOD
GOOD
of
W-TEST
,
I-TEST
NG
HIGH
LOW
of
W-TEST
,
I-TEST
READY
READY
lamp
REMOTE
REMOTE
lamp
POWER ON
----------
Finish of setting
Press
ENTER
key, then the tester returns to READY status, memorizing the set having
been made.
When the
EXIT
key (
SHIFT
and
ENTER
at a time) is pressed, the setting mode for
the condition of status output is interrupted and the tester becomes READY status.
The setting mode for the condition of status output in this case is that before entering the
setting mode for the condition of status output
ON/OFF
ENTER
WRITE
READ
LOCK
○
○
BUZZER
W
READY
REMOTE
DISCHARGE
KEY LOCK
EXIT
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
SHIFT
M
Ω
LOW
GOOD
HIGH
LOW
GOOD
HIGH
LOW
○
W-TEST
○
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
HIGH
mA
○
○
○
○
I
I-TEST
I
I-W
W-I
◎
◎
VOLTAGE SLIDER
AC TEST VOLTAGE
CURRENT/RESISTANCE
kV