18
5. Setting items in each mode
5.1
●
READY status
When turned ON the
POWER
switch
①
, the lamp test is done and then
READY
switch is lit up, entering into READY status.
In READY status of automatic test mode (W-I, I-W), the setting of test condition of the
withstanding voltage test and insulation resistance test are alternatively displayed at the
cycle of 2 seconds.
The test condition when the power was shutdown last time is displayed.
Pressing the
START
switch
③
starts the test.
In READY mode, the setting of the following 5 items can be done.
(1) Test condition
Refer to the article 8~9 (P22~37)
(2) Key lock
Refer to the article 12 (P46)
(3) Buzzer sounding
Refer to the article 18 (P59)
(4) Status output conditions Refer to the article 16.3 (P56)
(5) Special test mode
Refer to the article 13 (P47)
①
Double action
②
GOOD hold
③
Momentary
④
FAIL mode
5.2
●
Setting mode of test condition
In READY status, by pressing the
(or
) key,
READY
is turned off and the
tester enters into the test condition setting mode.
In the test condition setting mode, the test mode and condition can be set or changed.
A press of
ENTER
key finishes the setting and the tester becomes READY status.
Item to set
Test mode
Test condition
Ref. art. 8~9 (P22~37)
5.3
●
Memory write-in mode
After setting the test condition in the test condition setting mode, press the
WRITE
key
(
SHIFT
+
at a time), then the memory number blinks, being ready to write in the
memory.
In the memory write-in mode, 9 memory sets can be written. Each memory set consists
of 12 items of test conditions which are set in the test condition setting mode.
A press of
ENTER
key finishes the setting and the tester becomes READY status.
Item to set
Test condition
→
Memory write-in (No.1)
→
Test condition
→
Ref. art. 8~9 (P22~37) Ref. art. 10.2 (P38) Ref. art. 8~9 (P22~37)
Memory write-in (No.2)
Test condition
→
Memory write-in (No.9)
Ref. art. 10.2 (P38) Ref. art. 8~9 (P22~37) Ref. art. 10.2 (P38)
5.4
●
Memory read-out mode
In READY status, by pressing the
READ
key (
SHIFT
and
at a time)
, a memory
No. blinks and the tester becomes ready to read out the memory. In the memory read out
mode, one of the max. 9 memories written in [ref. art. 10.2 (P38)] can be called up and
read out.
A press of
ENTER
key finishes the setting and the tester becomes READY
status.
Item to set
Memory read out
(Select memory No.)
Ref. art. 10.3 (P39)