7
Explanation of communication method
3.2
●
Basic format of read-out command
When the “?” is added to the command letters sent from the host, 8527 transacts it as
read-out command. To the read-out command, 8527 adds “
=
parameter ” to the command
letters and transmits it to the host.
Command from the host side
: Command letters?
Response from 8527 to the host
: Command letters=parameter
In case of error, 8527 transmits the error code to the host.
Refer to the
article 6 (P34) Error
.
Example
3.6
Command
:
ILOW?
Reads out the low limit value of
resistance
of
insulation
resistance
test.
Response
is
:
ILOW=10.0MOHM
3.3
●
Basic format of setting and operation
○
When the “
=
” is added to the letters of setting command from the host side, 8527
transacts it as setting command.
○
“
=
” is not necessary for the operation command START and RESET.
Setting command from the host side
:
Command letters=
Operating command from the host side :
Command letters
Example 3.7 In case of setting command
Effective
command
:
MODE=WI
・・・・・
Test mode is set to W-I
Response
:
ERROR=0
・・・・・・
When Response Setting is ON.
Response
:
No
response
・・・・・・・・
When Response Setting is OFF.
In case of error, the error code is transmitted to the host.
Example 3.8 In case of operation command
Effective
command
:
In
case
of
START
Starts the test.
Response
:
ERROR=0
・・・・・・
When Response Setting is ON.
Response
:
No
response
・・・・・・・・・
When Response Setting is OFF.
In case of error, the error code is transmitted to the host.