22
Explanation of command
4.2.27
IHIGH=
(setting of high limit of resistance value of insulation resistance test)
Function
Makes the setting of high limit of resistance value of insulation
resistance test.
Structure
IHIGH=
High limit of resistance value
High limit of resistance value OFF or 0.2MOHM~2000MOHM is to
be set.
Note
: High limit value of resistance can not be lower than the low
limit value of resistance.
Transmission
IHIGH=100.0MOHM
…. Sets the high limit value of resistance of insulation
resistance test at 100.0M
Ω
.
Response
When 8527 received the effective command setting.
ERROR=0
…….…….... When the Response Setting is ON.
No response ………………. When the Response Setting is OFF.
4.2.28
IHIGH?
(read-out of high limit of resistance value of insulation resistance test)
Function
Reads out the high limit of resistance value of insulation resistance test.
Structure
IHIGH?
Transmission
IHIGH?
Response
IHIGH=100.0MOHM
…. Indicates the high limit value of resistance of insulation
resistance test 100.0M
Ω
.