13
Explanation of command
4.2.9
MODE=
(setting of test mode)
Function
Makes the setting of test mode.
Structure
MODE=
Parameter
Parameter
WI
: Automatic sequential test mode,
withstanding
voltage
test
→
insulation resistance test.
IW
: Automatic sequential test mode,
insulation
resistance
test
→
withstanding voltage test.
W
: Single test mode of withstanding voltage test
I
: Single test mode of insulation resistance test
Transmission
MODE=WI
Test mode is set to automatic sequential test mode of WI (withstanding voltage test
→
insulation resistance test).
Response
When 8527 received the effective command setting.
ERROR=0
…….………. When the Response Setting is ON.
No response ……………….. When the Response Setting is OFF.
4.2.10
MODE?
(read-out of test mode)
Function
Reads out the test mode being set.
Structure
MODE?
Transmission
MODE?
Response
MODE=I
……………… When the test mode setting is I, insulation resistance test.