31
Setting of test condition for insulation resistance test
9.2
●
High limit of resistance value
Adjustable range: 0.2M
Ω
~2000M
Ω
, OFF
Note-1
: The adjustable range is 0.2~9.9M
Ω
(resolution 0.1M
Ω
) and 10~2000M
Ω
(resolution 1M
Ω
).
Note-2
: The high limit of resistance value can not be lower than that of low limit, so
please apply either corrective solution below:
1. When the low limit value is determined, set the high limit value to exceed the
value of low limit, or turn OFF the setting.
2. When the high limit value is determined, set the low limit value not to exceed
the value of high limit.
Note-3
: When the setting is turned OFF, no judgement for the high limit is made.
When the setting is restored from OFF, and when the high limit value is lower
than the low limit value, the high limit value of resistance is replaced with
2000M
Ω
.
[When turning OFF the setting of high limit of resistance value]
To enter setting mode
①
In READY status, press
or
key, then the test mode lamp blinks.
Test mode lamp moves up and down with
▲
or
▼
key. Make the required test
mode lamp blinking (
W-I
,
I-W
or
I
lamp).
②
Press
or
key and make the test mode lamp lit.
To turn OFF the setting of high limit of resistance value
①
Press
or
key and select the status that the current/resistance display blinks,
the
M
Ω
is lit up and the
HIGH SET
is also lit.
②
Next, press
ON/OFF
key and select the status that the display blinks with
(refer to the above figure).
To move to the previous setting
Press
key, then changes to the
setting of test range of insulation resistance test
.
To the next setting
Press
key, then changes to the
setting of low limit of resistance value
.
Finish of setting
Press
ENTER
key, then the tester returns to READY status, memorizing the set having
been made.
The slider moves up or down when the test mode is W-I or I-W.
When the
EXIT
key (
SHIFT
and
ENTER
at a time) is pressed in the setting mode,
the setting mode for the test condition is interrupted and the tester becomes READY
status.
The test conditions in this case are those before entering the setting mode of test condition.
ON/OFF
ENTER
WRITE
READ
LOCK
○
BUZZER
W
READY
REMOTE
DISCHARGE
KEY LOCK
EXIT
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
SHIFT
LOW
GOOD
HIGH
LOW
GOOD
HIGH
LOW
○
W-TEST
○
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
mA
○
○
I
I-TEST
I
I-W
W-I
HIGH
M
Ω
○
●
○
●
●
VOLTAGE SLIDER