DP4500/3500 TROUBLESHOOTING
5 - 114
December 2000 © TOSHIBA TEC
Defective area
Step
Check Items
Prescription
Developer material
Cleaning roller
Fuser roller
Cleaning blade
Toner recovery auger
1
Has the specified developer material been
used?
2
Is the cleaning roller damaged or has it
reached its PM life?
3
Are there bubble-like scratches on the
fuser roller (125mm pitch on the copied
image)?
4
Has the fuser roller reached its PM life?
5
Is the pressurization of the fuser roller
normal?
6
Is the temperature of the fuser roller
normal?
7
Is the blade in proper contact with the
drum?
8
Has the cleaning blade been turned up?
9
Is the toner recovered normally?
Use the specified developer material and toner.
Replace the roller.
Replace the fuser roller. Check and adjust the tem-
perature control circuit.
Replace it.
Check and adjust the pressurization mechanism.
Check the adjustment values of fuser roller
temperature?
08-410:12 (200
°
C)
08-411:12 (200
°
C)
Check the blade.
Replace the blade.
Check and replace drum if necessary.
Clean the toner recovery auger.
Check the pressure of the cleaning blade.
(20) Poor cleaning
Feeding direction
Summary of Contents for DP3500
Page 91: ...DP4500 3500 PREVENTIVE MAINTENANCE 3 12 December 2000 TOSHIBA TEC ...
Page 95: ...DP4500 3500 SUPPLIES 4 4 December 2000 TOSHIBA TEC ...
Page 213: ...DP4500 3500 TROUBLESHOOTING 5 118 December 2000 TOSHIBA TEC ...
Page 253: ...DP4500 3500 UPDATING THE FIRMWARE 6 40 December 2000 TOSHIBA TEC ...
Page 256: ...DP4500 3500 WIRE HARNESS CONNECTION DIAGRAMS 7 4 December 2000 TOSHIBA TEC ...