35
Operating Instructions TNI
®
20s oxy Version 2.7
Version 1.6.3 July 2010
rule: 5 ml disinfectant Wofasteril
®
per liter
disinfection soluton.
6.2.2 Cleaning and preparation interval
A conceptual distinction is made here between
cleaning
, which
should normally be done before or after every use, and the
preparation
work, consisting of pre-cleaning and disinfection/
sterilization, which is carried out at greater time intervals or when
there is a change of patient.
The following table contains the intervals for cleaning and disinfection
or sterilization. It is assumed that the patient uses the device
daily.
Frame sur-
faces
Humidifier
cover
Humidifier
basin
Water
reservoir
Cleaning
interval
Per patient/
daily
Per patient/
daily
Per patient/
daily
Per patient/
daily
Disinfec-
tion
interval
Per patient/
weekly
Per patient/
weekly
Per patient/
weekly
Per patient/
weekly
Steriliza-
tion
interval
not applicable not applicable not applicable Per patient/
weekly
Separator pa-
nels/air vents
Connecting
hose
Connection
units
Applicator
Cleaning
interval
Per patient/
weekly
Per patient/
weekly
Per patient/
weekly
Per patient/
daily
Disinfec-
tion
interval
Per patient/
weekly
Per patient/
weekly
Per patient/
weekly
Per patient/
weekly
Steriliza-
tion
interval
not applicable
not applicable not appli-
cable
not appli-
cable
Always wear protective glasses and
disposable gloves while working with the
disinfectant solution!
IMPORTANT!
6 Hygiene