41
Operating Instructions TNI
®
20s oxy Version 2.7
Version 1.6.3 July 2010
cover and separator panels from the disinfectant bath and
proceed with the steps specified in chapter 6.2.4.4.
6.2.4.3.3 Connecting hose (immersion disinfection)
· Immerse the connecting hose in the second immersion
basin in a bath of disinfectant and leave there for the
application time specified in chapter 6.2.1.1.1 „Immersion
disinfection“. Make sure that all cavities are completely filled
with disinfectant and that there are no air bubbles in the
tube.
· After expiry of the application time, remove the connecting
hose from the disinfectant bath and proceed with the steps
specified in chapter 6.2.4.4.
6.2.4.3.4 Applicator (immersion disinfection)
· Immerse the applicator with the nose prongs first up to the
Y-piece (junction of both thinner tubes in the long part of
the tube) in the third immersion basin sufficiently filled with
disinfection solution.
· Then fill the applicator with disinfectant by slowly pressing
the 60-ml syringe via the tube opening in the applicator plug,
which is outside the immersion basin, until the applicator is
completely filled.
. Now place the applicator in the third immersion basin for the
application time specified in chapter 6.2.1.1.1 „Immersion
disinfection“. Make sure that the applicator plug lies outside
of the immersion basin, so that it does not come into contact
with the disinfectant.
· After expiry of the application time, take the applicator out of
the disinfectant bath and let the entire quantity of disinfectant
drain out of the applicator (hold the applicator up with the
plug on top) and proceed with the steps specified in chapter
6.2.4.4.
6.2.4.3.5 Applicator plug (spray / wipe disinfection)
· Spray the outer surface of the cleaned applicator plug
with 70% ethyl alcohol and wipe it thoroughly with a cloth
moistened with 70% ethyl alcohol.
6 Hygiene