8 Error Codes
115
The error may have
the following causes:
External heat source
is affecting the ambi-
ent temperature sen-
sor.
Operation outside
the permitted
ambient conditions.
Internal error
The sensor is located at the
connecting lead between the
blower and humidifier units at the
back of the humidifier unit. Ensure
adequate distance from other
objects, especially from external
heat sources (heating, radiators,
etc.) and adequate air circulation.
Note the permitted ambient con-
ditions (see chapter 7 „Technical
Specifications“).
Contact the Service Center.
116
The error may have
the following causes:
Operation outside
the permitted
ambient conditions.
Internal error
Note the permitted ambient con-
ditions (see Chapter 7“Technical
Specifications“).
Contact the Service Center.
120
he error may have
the following causes:
Water has seeped
into the space (at the
air inlet side) behind
the separator panels
and has penetrated
the humidifier unit.
System is leaking:
Connecting points on
blower side are de-
fective.
Switch off TNI
®
20 oxy via the
master switch („0“ position) at the
back of the blower. Disconnect the
mains lead from the power outlet.
Contact the Service Center.
Switch TNI
®
20 oxy to standby
mode. Allow the unit to cool down.
Check the connecting points of the
humidifier unit to the connecting
hose.