24
Operating Instructions TNI
®
20s oxy Version 2.7
Version 1.6.3 July 2010
4.6 Set the prescribed amount of oxygen
- Oxygen may only be added for the product
variant TNI
®
20s oxy.
- The oxygen flow meter at the connection
unit may only be operated in an upright ver-
tical position (see Chapter 4.1 Assembling
and setting up the TNI
®
device)
- Before you fasten the applicator to the
patient, first set the prescribed air-oxygen
mixture at the setting valves of the connec-
tion unit.
- Check the oxygen flow value at regular
intervals.
- Determine the exact oxygen concentration
in the air-oxygen mixture with an indepen-
dent measuring device.
Fire hazard when using oxygen!
IMPORTANT!
WARNING!
- Check before opening the setting valve
whether the float is on
the floor of the metering tube.
- Set the desired amount of oxygen at the oxygen flow meter by
turning the knob at the setting valve
as follows:
increase oxygen flow:
turn knob counterclockwise
decrease oxygen flow:
turn knob clockwise
4 Startup
- The current oxygen flow value is determined
by the float (ball) in a metering tube. Read the
value on the upper edge of the float.
- Note that a maximum oxygen flow of 16 l/min
may be set on the oxygen flow meter.