8 Error Codes
Reservoir is damaged
(e.g. dented).
Internal error
Switch the TNI
®
20 oxy to standby
mode. Allow the unit to cool down.
Check the reservoir for any visual
damage. If damaged, use a new
reservoir.
Contact the Service Center.
131
Internal error
Contact the Service Center.
132
The error may have
the following causes:
Applicator is not con-
nected.
Heating filament in
applicator is broken:
replacement interval
exceeded.
Internal error
Connect the applicator to the front
side of the humidifier unit.
Observe the cleaning and replace-
ment intervals (see chapter 6 „Hy-
giene“). Replace the applicator
tube, if necessary.
Contact the Service Center.
133
Internal error
Contact the Service Center.
134
Internal error
Contact the Service Center.
140
The error may have
the following causes:
Communication
between blower unit
and humidifier unit
was disrupted.
Internal error
Switch off TNI
®
20 oxy via the mas-
ter switch („0“ position) at the back
of the blower. Check for proper
connection of the the connecting
cable between both units.
Contact the Service Center.
220
221
230
Blower unit error
Switch off TNI
®
20 oxy via the mas-
ter switch („0“ position) at the back
of the blower. Allow the unit to
cool down. Restart the device. If
the error message appears again
as soon as the device is switched
on, please contact the Service
Center.