8 Error Codes
System is leaking:
the connecting hose
is not properly con-
nected, has slipped
off or has a leak.
Applicator tube is
squeezed together
or has a kink.
Air flow is blocked.
Water has seeped
into the hose outputs
behind the separator
panels. Flow sensor
is defective
System is leaking:
The oxygen and
safety tube is not
proberly connected,
has slipped off or
has a leak.
Internal error
Switch TNI
®
20 oxy to standby
mode. Allow the unit to cool down.
Check the safe and tight connection
of the connecting hose between
the blower unit and the humidifier
unit. If a leak is detected, the con-
necting hose must be replaced!
Use a new applicator. Place the
applicator tube on top of the bed
covers during therapy.
TNI
®
must not be used when the
upper airways are completely
closed.
Switch off TNI
®
20 oxy via the
master switch („0“ position) at the
back of the blower. Disconnect the
mains lead from the power outlet.
Contact the Service Center.
Switch TNI
®
20 oxy to standby
mode. Allow the unit to cool
down. Check the safe and tight
connection of the oxygen and
safety tube between the blower
unit and the oxygen source. If a
leak is detected, the tube must be
replaced!
Contact ther Service Center.
122
The error may have
the following causes:
System is leaking.
Connecting points on
the applicator side
are defective.
System is leaking.
Housing lid or hu-
midifier cover are not
sealing properly.
Switch TNI
®
20 oxy to standby
mode. Allow the unit to cool down.
Check the connecting points of the
humidifier unit to the applicator.
Switch TNI
®
20oxy to standby mode.
Allow the unit to cool down. Check
the correct position of both covers.
Check that the seal on the humidi-
fier unit‘s cover is not defective.