39
Operating Instructions TNI
®
20s oxy Version 2.7
Version 1.6.3 July 2010
Subsequently thoroughly rinse all components with distilled
water.
· Wipe the humidifier basin out with a lint-free cloth moistened
with clear drinking water. Then wipe the humidifier basin out
with a lint-free cloth moistened with cleaning solvent. The
separator panel chambers with air inlet and outlet must be
thoroughly cleaned. The cleaning procedure can be made
more effective by using a cotton swab soaked in cleaning
agent to clean the corners of the separator panels and the
air inlet and outlet.
· Pre-clean the connecting hose under running water,
immerse it in cleaning solvent and wipe the outer surface
with a cloth soaked in cleaning solvent. Then thoroughly
rinse the connecting hose under distilled water (inside and
outside).
· After this cleaning procedure, wipe the device surfaces,
humidifier basin, the separator panels, the humidifier cover
and the reservoir with a dry cloth in order to remove all
residual moisture. Be sure not to miss the corners behind
the separator wall panels. Hang the connecting hose up
vertically to dry.
· Pre-clean the applicator under running water and wipe off
the outside of the applicator with a lint-free cloth moistened
with cleaning solvent. If there are any fluids or secretions in
the applicator (e.g. due to sneezing), proceed as follows: Fill
the reservoir with drinking water or sterile water up to the
filling mark and place it in the device. Connect the applicator
to the humidifier unit, adjust an air flow of 20 l/min (only AIR)
and switch the device to operating mode. Press the Standby
button once to switch to the drying mode. The device
automatically finishes the drying phase after a minimum of
30 minutes and maximum 90 minutes, depending on the
ambient temperature. Any residual liquid secretions will be
blown out over the nose prongs during the drying process.
Remove this and then wipe the outside of the applicator
with a cloth soaked in biodegradable cleaning solvent. Dry
it well in order to avoid calcareous deposits (lime).
Thoroughly clean all the preparation accessories after
completion of cleaning!
6 Hygiene