6 Hygiene
Clean the preparation accessories
thoroughly after completing the disinfection
procedure! Use a fresh pair of disposable
gloves for the following steps!
IMPORTANT!
6.2.4.4 Rinsing and drying
6.2.4.4.1 Humidifier basin
· To avoid direct or indirect contact of disinfectant residues
with the patient, after the application time thoroughly wipe
the humidifier basin with a clean lint-free cloth (cellulose/
cosmetic tissue) and distilled water. It is best to clean out
the corners of the separator panel chambers and the air
inlet and outlet with a cotton swab soaked in distilled water.
Rinsing removes the components of the disinfectant which
are detrimental to health.
· Subsequently wipe the humidifier basin dry with a lint-free
cloth.
6.2.4.4.2 Humidifier cover and separator panels
· Also likewise rinse the humidifier cover and separator wall
panels with distilled water and wipe the humidifier basin dry
with a lint-free cloth.
6.2.4.4.3 Connecting hose
· Rinse the connecting hose inside and out with distilled water
and hang vertically to dry.
6.2.4.4.4 Applicator
· Immerse the applicator with the nose prongs first up to the
Y-pieces (junction of both thinner tubes) in distilled water.
· Then fill the applicator with distilled water by slowly pressing
the 60-ml syringe via the tube opening in the applicator plug,
until the applicator is completely filled.
· Afterwards, force the water out of the applicator again by
pressing the air-filled syringe.
· Dry the applicator on the outside (especially the applicator
plug) with a lint-free, soft cloth.