115
SC-F70
SC-F70
PID Operation Operating Instructions
9.11 General Specifications
(1) Insulation resistance:
DC 500 V, 20 M
W
minimum
between measurement terminals and
ground terminals
DC 500 V, 20 M
W
minimum
between power terminals and
ground terminals
(2) Dielectric strength:
AC 1000 V for one minute between
measurement terminals and ground
terminals
AC 1500 V for one minute between
power terminals and ground
terminals
(3) Line voltage:
AC 90 V to AC 264 V including
line voltage fluctuations, 50/60 Hz,
rated AC 100V to AC 240 V
(4) Power consumption:
13 VA at AC 240V
10 VA at AC 100V
(5) Effect of power outage:
No effect on operation if power
outage is 50 msec. or less
(6) Warm-up time:
1 hour
(7) Memory backup:
Data is backed up by a lithium
battery.
Battery service life: Approximately
10 years, depends on product
storage time, storage environment,
usage, and other conditions
(8) Weight:
Approximately 500 g (1.1 lb)
(9) Accessories:
A pair of mounting brackets
9.12 Environmental Conditions (Normal
Operation)
(1) Ambient temperature:
0 to 50˚C (32 to 122˚F)