8 Specifications
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
8-9
Desflurane:
0
– 1 Vol% ± 0.15 Vol%
1
– 5 Vol%
± 0.2 Vol%
5
– 10 Vol% ± 0.4 Vol%
10
– 15 Vol% ± 0.6 Vol%
15
– 18 Vol% ± 1.0 Vol%
> 18 Vol% unspecified
8.7 MIRUS Controller Monitoring System
Safety monitoring
Patient:
Breathing activity
Alarms:
Apnoea
Inspiratory tidal volume:
Low Vti
Agent dosage monitoring
Dosage control:
Agent
Dosage start:
Start of inspiration.
Dosage end:
End of inspiration or end of delivered anaesthetic agent
volume.
Control:
Monitoring of applied volume of anaesthetic agent.
Comparison of calculated with measured volume.
Alarms:
Dosage failed! No delivery.
Gas monitoring
Agent:
Isoflurane
Sevoflurane
Desflurane
Numerical data:
etVA (end expiratory volatile agent concentration)
Alarms:
Low etVA, High etVA
Metabolic gases:
CO
2
Numerical data:
etCO
2
(end expiratory CO
2
concentration)
Alarms:
Low etCO
2
, High etCO
2