7 Alarms and messages
7-2
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
7.1 Alarm modality
The MIRUS Controller distinguishes between high and low priority alarms. The following
describes the MIRUS Controller’s high and low priority alarms.
Alarms only occur during operation and are shown in the following way:
•
Alarm lights:
red or yellow alarm lights activated
•
On screen:
alarm bar with text
+
red or yellow highlighted icon
(e.g. low etCO
2
high priority alarm).
High priority
Activity
Screen
Red alarm light
Audible alarm
Alarm is active.
- Display alarm message in
black characters on red
background.
- Start timer.
ON, flashing
ON
Confirming the
alarm by activating
the alarm silence
button.
- Display alarm message in
black characters on red
background.
- Timer
- Start “2 min” silence.
ON, continuous
paused
Alarm is still active
after the “2 min”
silence time.
- Display alarm message in
black characters on red
background.
- Timer
ON, flashing
ON
Low priority
Activity
Screen
Yellow alarm light Audible alarm
Alarm is active.
- Display alarm message in
black characters on yellow
background.
ON, continuous
ON
Confirming the
alarm by activating
the alarm silence
button.
- Display alarm message in
black characters on yellow
background.
ON, continuous
OFF
since 2 sec
Low et CO
2 [11]
et Sevo: 1.9
Vol%
et CO
2
: 25
mmHg