2 Safety
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
2-11
2.4 Safety features
The MIRUS Controller is equipped with monitor functions for monitoring the device status and
thus detecting changes in parameters. Changes in parameters may be caused by:
•
Changes in patient status
•
Changes in the settings
•
Adjustment and/or operating errors
•
Device malfunctions
•
Changes in the electrical supply
•
Changes in the anaesthetic gas supply
The following safety functions are integrated in the MIRUS Controller to warn the operator if
the patient is at risk.
Device alarms
Control
Monitoring
Anaesthetic supply
Internal reservoir empty, Dosage failure
Electrical supply
Failure of mains power supply
Failure of emergency battery
Device
Watchdog alarm
Incorrect positioning of the MIRUS Controller (inclination angle
> 25°)
Interface
Occlusion or disconnection of Interface
MIRUS Reflector
Service hours of the MIRUS Reflector
Comment:
The MIRUS Reflector control is only active when there is a continuous connection between the
interface connector and the MIRUS Controller (mains power supply is connected). If the
interface connector is disconnected or the MIRUS Controller is switched off (mains power
supply not connected), the timer of the MIRUS Reflector control is reset.
Patient alarms
Control
Monitoring
Anaesthetic gas
concentration
Low and High etSevo / etIso / etDes
CO
2
concentration
Low and High etCO
2
Apnoea
Apnoea alarm