7 Alarms and messages
7-18
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
No.
Message
(in order of
testing)
Set condition
Required user
action
Device action
B06
System test failed!
Reflector not in
Park Bay. Please
insert the
Reflector and
repeat the System
test.
MIRUS Interface
detected.
MIRUS Reflector
not detected in
Park Bay.
Insert MIRUS
Reflector into
Park Bay
(recommended)
or turn off MIRUS
Controller via turn
off button.
Stops System test
sequence until
user activity or
performs turn OFF
sequence after
user selection.
B07
System test failed!
Please adjust the
position of the
Reflector in the
Park Bay and
repeat the System
test.
Leakage detected. Check position of
MIRUS Reflector
in Park Bay or turn
off MIRUS
Controller via turn
off button.
Stops System test
sequence until
user activity is
done. Then
repeats test step.
Or performs turn
OFF sequence
after user
selection.
B08
System test failed!
Please remove the
Interface plug,
insert it again and
repeat the System
test.
Leakage detected. Check
MIRUS Interface
connection to
MIRUS Controller
or turn off MIRUS
Controller via turn
off button..
Stops System test
sequence until
user activity is
done. Then
repeats test step.
Or performs turn
OFF sequence
after user
selection.
B09
System test failed!
Please exchange
the Reflector and
repeat the System
test.
Leakage detected. Exchange MIRUS
Reflector and
repeat the System
test or turn off
MIRUS Controller
via turn off button.
Stops System test
sequence until
user activity is
done. Then
repeats test step
or performs turn
OFF sequence
after user
selection.
B10
Thank you.
Requested user
activity performed.
Perform request.
Continues test.
B11
The device is
inclined. Please
position device on
a level surface.
MIRUS Controller
inclined >25° for >
30 sec.
Move MIRUS
Controller to level
surface or turn off
via turn off button..
Alarm or performs
turn OFF
sequence after
user selection.
B12
Mains supply lost!
Restore mains
supply or the
device will turn off
automatically.
AC lost.
Connect AC line
supply or turn off
MIRUS Controller
via turn off button.
Operates on
battery. Performs
turn OFF
sequence after
user selection or if
AC isn't restored
within 30 seconds,
the system shall
shut down into
OFF state.
Summary of Contents for Mirus MC DES
Page 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Page 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Page 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...