7 Alarms and messages
7-14
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
7.4 Other messages
The MIRUS Controller software includes self-tests that determine if the operating software is
functioning properly and if the electronic and pneumatic circuits are functional.
The self-tests include:
•
Power up test (PUT)
•
System test (ST)
7.4.1 Messages during Power up test
Messages generated during the Power up test are shown on the screen as follows:
No.
Message
(in order of
testing)
Set condition
Required user
action
Device action
A01
One moment
please...
Micro controller
system check
successful.
None.
System performs
automatic test
procedure.
A02
The agent
reservoir is empty.
To continue
please refill the
device.
To open Fill port,
please confirm.
Internal reservoir
fill level < 75 mL.
Refill the MIRUS
Controller or turn
off via turn off
button.
Waits for refill or
performs turn OFF
sequence after
user selection.
A03
Thank you!
Please start the
filling process.
0 mL VA filled.
Internal reservoir
fill level < 190 mL.
The fill port flap is
open for < 4 min.
Refill the MIRUS
Controller.
Shows changing
filling level.
A04
The maximum
level is reached.
Please close the
Fill port.
Internal reservoir
fill level 190 mL.
Close Fill port flap
or turn off MIRUS
Controller via turn
off button.
Waits for closed
Fill port flap or
performs turn OFF
sequence after
user selection.
HW 2.n
SW 2.03.02
The agent reservoir is empty.
To continue please refill the device.
To open Fill port please confirm.