8 Specifications
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
8-7
End tidal anaesthetic concentration
etISO min:
Range:
0.0 to 5.0 Vol%
Increment:
0.1 Vol% @ single tap on the touch button
0.5 Vol% @ pressing the touch button for > 1 sec
etISO max:
Range:
0.5 to 5.0 Vol%
Increment:
0.1 Vol% @ single tap on the touch button
0.5 Vol% @ pressing the touch button for > 1 sec
etSEVO min:
Range:
0.0 to 8.0 Vol%
Increment:
0.1 Vol% @ single tap on the touch button
0.5 Vol% @ pressing the touch button for > 1 sec
etSEVO max:
Range:
0.5 to 8.0 Vol%
Increment:
0.1 Vol% @ single tap on the touch button
0.5 Vol% @ pressing the touch button for > 1 sec
etDES min:
Range:
0.0 to 12.0 Vol%
Increment:
0.1 Vol% @ single tap on the touch button
0.5 Vol% @ pressing the touch button for > 1 sec
etDES max:
Range:
0.5 to 12.0 Vol%
Increment:
0.1 Vol% @ single tap on the touch button
0.5 Vol% @ pressing the touch button for > 1 sec
Apnoea time:
Range:
15 to 60 sec
Increment:
5 seconds
Alarm volume:
Range:
50 to 100%
Increment:
10% @ single tap on the touch button
Patient data settings
Age:
Range:
10 to 115 years
Increment:
1 year @ single tap on the touch button
5 years @ pressing the touch button for > 1 sec
Size:
Range:
100 to 250 cm
Increment:
5 cm @ single tap on the touch button
10 cm @ pressing the touch button for > 1 sec
Weight:
Range:
15 to 125 kg (ideal body weight)
Increment:
1 kg @ single tap on the touch button
5 kg @ pressing the touch button for > 1 sec