Observation
Possible cause
Recommended action
Verification failed but plate is
undamaged
(continued)
The plate was improperly
prepared.
Repeat the verification with a new properly
prepared plate.
Note:
Each RNase P plate can only be used
once.
Open the data file for verification in the desktop
software or the Connect platform to view the
flags and troubleshooting details for failed wells.
If the verification fails again, order a replacement
plate. Use Smart Help if the replacement plate
fails (see “Request technical support with Smart
Help” on page 159).
Chapter 7
Calibrate and verify instrument performance
Perform verification using RNase P plates
7
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide
97