Operation
Service Menu
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
3-69
Lamp Setting
The Lamp Setting screen is used to adjust the detector intensities. The
display shows the intensities of Cell A and Cell B. The next line of the
display shows the current lamp setting. Adjust the lamp setting until the
intensities are about 100 kHz.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Lamp Setting
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save the new lamp setting.
Detector Calibration
The Detector Calibration screen is used to set the calibration factors for
detector A and detector B. Calibration factors are computed so that both
detectors read 100,000 Hz on zero air. If either detector reads lower than
75,000 Hz or higher than 125,000 Hz, this calibration will not be
successful. It is crucial that the lamp intensity be set such that both detector
intensities fall inside this range before performing this calibration.
PUMP PRESSURE: 753.4 mmHg
0.010 LPM
TESTING, WAIT 20 SECS
RANGE AVG DIAGS ALARM
BENCH LAMP SETTING:
CELL A INT: 98425 Hz
CELL B INT: 97465 Hz
LAMP SETTING: 72.9 %
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM