Operation
Instrument Controls Menu
3-46
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Units
The Units screen allows the user to enter the units for the selected analog
input channel. The units are displayed on the diagnostic screen and in
datalogging and streaming data. The units may be from 1 to 3 characters in
length, and defaults to V (volts).
•
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel >
Units
.
•
Press
to save the new value.
Decimal Places
The Decimal Places screen allows the user to select how many digits are
displayed to the right of the decimal, from 0 to 6, with a default of 2.
•
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel >
Decimal Places
.
•
Use
and
to increment or decrement the value.
•
Press
to save the new value.
Number of Table Points
The Number of Table Points screen allows the user to select how many
points are used in the conversion table. The points range from 2 to 10, with
a default of 2.
ANALOG INPUT UNITS:
CURRENTLY: V
V
SS
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
DECIMAL PLACES:
CURRENTLY: 2
SET TO: 3 ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM