Operation
Service Menu
3-74
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Restore Default Pressure
Calibration
The Restore Default Calibration screen allows the user to reset the pressure
calibration configuration values to factory defaults.
•
In the Main Menu, choose Service > Pressure Calibration >
Set
Defaults
.
•
Press
to warn user and enable restore with
.
•
Use
to overwrite pressure sensor calibration parameters with
factory default values when pressed after
.
Flow A and B Calibration
The Flow A Calibration menu is used to calibrate the flow A sensor to zero,
span, or restore factory default values. The flow A calibration is visible only
when the calibrator is in service mode. For more information on the service
mode, see “Service Mode” earlier in the chapter. The Flow B Calibration
menu functions the same way.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Flow Calibration A
or
B.
RESTORE DEFAULT CAL:
RESTORE DEFAULT CAL:
RESTORE
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
FLOW SENSOR CAL:
>ZERO 200
SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM