Servicing
Front Panel Board Replacement
6-22
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
2. Unplug all connectors. Note the locations of the connectors to facilitate
reconnection.
3. Remove the two retaining screws from the top of the board.
4. Pop off the measurement interface board from the two mounting studs
and remove the board.
Figure 6-9.
Replacing the Measurement Interface Board
5. To install the measurement interface board, follow previous steps in
reverse.
6. Re-install the measurement bench. Refer to “Removing the
Measurement Bench and Lowering the Partition Panel” in this chapter.
7. Calibrate the pressure transducer, flow transducers, and bench
temperature sensor as described earlier in this chapter.
Front Panel Board
Replacement
Use the following procedure to replace the front panel board (
Equipment Required:
Photometer Board
Ozonator Board
Measurement Interface Board