Operation
Service Menu
3-78
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Analog Output Calibrate Zero
The Analog Output Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog output. The operator must connect a meter
to the output and adjust the output until it reads 0.0 V on the meter.
•
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel >
Calibrate Zero
.
•
Use the
and
to increment or decrement the numeric
value.
•
Press
to save the value.
Analog Output Calibrate
Full-Scale
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
a meter to the output and adjust output until it reads the value shown in the
set output to: field.
•
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel >
Calibrate Full Scale
.
•
Use the
and
to increment or decrement the numeric
value.
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CAL: ZERO
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 100
SAVE VALUE
©ª
INC/DEC
SET OUTPUT TO: 0.0 V
RANGE AVG DIAGS ALARM