Servicing
Photometer Lamp Voltage Adjustment
6-28
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
7. Carefully slide lamp out of bench and instrument case through access
hole.
8. Carefully slide new lamp into place until it bottoms. Loosely tighten the
Allen screw on clamp. Tighten two screws holding the clamp to the
optical bench. Pull the lamp out approximately 1/32-inch to 1/16-inch
to allow for expansion when the lamp warms up; tighten Allen screw on
the clamp.
9. Plug new photometer lamp cable into the LAMP connector on the
photometer board.
10. Slide the insulation back on the end of the lamp.
11. Replace access hole cover and secure with the two retaining screws.
12. Re-install the instrument cover.
13. Reconnect power and turn instrument ON.
14. After lamp has stabilized (approximately 15 minutes), adjust the lamp
voltage in the Service Mode menu until the output from each detector is
about 100 kHz.
Photometer Lamp
Voltage Adjustment
Use the following procedure to adjust the photometer lamp voltage until the
output from each detector is approximately 100 kHz.
Note
After turning power ON, wait for the lamp to stabilize (approximately
15 minutes) before beginning the following lamp voltage adjustment
procedure.
1. From the Main Menu, press
to scroll to Service > press
>
to scroll to
Lamp Setting
> and press
.
The Bench Lamp Setting screen appears.