Operation
Alarms Menu
3-58
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Contact Information
The Contact Information screen displays the customer service information.
•
In the Main Menu, choose Diagnostics >
Contact Information
.
Alarms Menu
The alarms menu displays a list of items that are monitored by the
calibrator. If the item being monitored goes outside the lower or upper
limit, the status of that item will go from “OK” to either “LOW” or
“HIGH”, respectively. If the alarm is not a level alarm, the status will go
from “OK” to “FAIL”. The number of alarms detected is displayed to
indicate how many alarms have occurred. If no alarms are detected, the
number zero is displayed.
To see the actual reading of an item and its minimum and maximum limits,
move the cursor to the item and press
.
Items displayed are determined by the options installed. The zero or span
check is visible only if the zero/span check or auto calibration options are
enabled. The O
3
lamp temperature and ozonator level 1, 2, 3, 4, and 5
checks are visible only if the ozonator option is installed. The motherboard
status, interface board status, and I/O expansion board status (if installed)
indicates that the power supplies are working and connections are
successful. There are no setting screens for these alarms.
•
In the Main Menu, choose
Alarms
.
CONTACT INFORMATION:
CALL CENTER: 508-520-0430
WEB: WWW.THERMO.COM