C-Link Protocol Commands
Diagnostics
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
B-17
Diagnostics
cell a int
cell b int
The example that follows reports that the lamp intensity in cell A is 98,425
Hz.
Send:
cell a int
Receive:
cell a int 98425 Hz
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, and positive 15. Each
voltage value is separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2 14.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
Datalogging
clr records
This command will clear all long and short records that have been saved.
Send:
clear records
Receive:
clear records ok
set clr lrecs
set clr srecs
These commands will clear only the long records or only the short records
that have been saved. The example that follows clears short records.
Send:
set clr srecs
Receive:
set clr srecs ok