Operation
Service Menu
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
3-67
Pressure Check
The Pressure Check menu is used to manually control the flow of reference
or sample gas through Cell B. This enables the pressure reading of Cell B,
with either sample or reference gas, to be determined. Pump pressure is used
to test the pump. Selecting any of these menu items will disturb the analog
outputs.
•
In the Main Menu, choose Service >
Pressure Check
.
Sample Pressure
The Sample Pressure screen (read only) displays the pressure of the sample
gas in Cell B.
•
In the Main Menu, choose Service > Pressure Check >
Sample
Pressure
.
SERVICE:
>PRESSURE CHECK
LAMP SETTING
DETECTOR CALIBRATION
INTENSITY CHECK
O3 RESPONSE COEFFICIENT
MANUAL O3 LAMP SETTING
PRESSURE CALIBRATION
ª
RANGE AVG DIAGS ALARM
FLOW A CALIBRATION
FLOW B CALIBRATION
TEMPERATURE CALIBRATION
ANALOG OUT CALIBRATION
ANALOG INPUT CALIBRATION
DISPLAY PIXEL TEST
RESTORE USER DEFAULTS