Introduction
Specifications
1-4
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
*In non condensing environments.
Weight
Approximately 35 lbs.
Analog outputs
6 voltage outputs; 0
–
100 mV, 1, 5, 10 V (user selectable), 5% of
full-scale over/under range, 12 bit resolution, user selectable for
measurement input
Digital outputs
1 power fail relay Form C, 10 digital relays Form A, user selectable
alarm output, relay logic, 100 mA @ 200 VDC
Digital inputs
16 digital inputs, user select programmable, TTL level, pulled high
Serial Ports
1 RS-232 or RS-485 with two connectors, baud rate 1200
–
115200,
data bits, parity, and stop bits, protocols: C-Link, MODBUS, and
streaming data (all user selectable)
Ethernet connection
RJ45 connector for 10Mbs Ethernet connection, static or dynamic
TCP/IP addressing
Ozonator
Output
0.025-1.000 ppm at 3-4 LPM
Response (to 98% or 5
ppb of final value,
whichever is greater)
1 minute
Stability
±
4 ppb or
±
1% of reading, whichever is greater
Photometer