Operation
Service Menu
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
3-73
Calibrate Pressure Zero
The Calibrate Pressure Zero screen calibrates the pressure sensor at zero
pressure.
Note
A vacuum pump must be connected to the pressure sensor before
performing the zero calibration.
•
In the Main Menu, choose Service > Pressure Calibration >
Zero
.
•
Press
to save the current pressure reading as the zero reading.
Calibrate Pressure Span
The Calibrate Pressure Span screen allows the user to view and set the
pressure sensor calibration span point.
Note
The plumbing going to the pressure sensor should be disconnected so
the sensor is reading ambient pressure before performing the span
calibration. The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before calibrating.
•
In the Main Menu, choose Service > Pressure Calibration >
Span
.
•
Use
,
,
and
to move and change the value.
•
Press
to save set to value as actual value.
CALIBRATE PRESSURE ZERO:
CURRENTLY: 753.4 mmHg
SET TO: 0.0 mmHg ?
CONNECT VACUUM PUMP AND
SAVE ZERO PRESSURE
RANGE AVG DIAGS ALARM
CALIBRATE PRESSURE SPAN:
CURRENTLY: 753.4 mmHg
SET TO: 760.
0
mmHg ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM