Operation
Diagnostics Menu
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
3-57
Set Analog Outputs
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-scale
voltage, zero sets the analog outputs to 0 volts, and normal operation. The
example below shows the selected output state “ALL” is set to normal.
•
In the Main Menu, choose Diagnostics > Test Analog Outputs >
ALL,
Voltage Channel 1-6,
or
Current Channel 1-6
.
•
Press
or
to set output.
•
Press
to reset to normal.
Instrument Configuration
The Instrument Configuration screen displays information on the hardware
configuration of the calibrator.
Note
If the calibrator is in service mode, pressing
on the item will
toggle it yes or no (with the exception of purchased options such as dilution
and auto calibration).
•
In the Main Menu, choose Diagnostics >
Instrument Configuration
.
•
Press
to toggle instrument configuration (in service mode only).
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL
©
SET TO FULL SCALE
ª
SET TO ZERO
§
RESET TO NORMAL
RANGE AVG DIAGS ALARM
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD YES
RANGE AVG DIAGS ALARM