Servicing
Analog Output Calibration
6-14
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Analog Output
Calibration
Use the following procedure to calibrate the analog outputs if a meter
reading in the “Analog Output Testing” procedure differed by more than
one percent or after replacing the optional I/O expansion board.
Equipment Required:
Multimeter
1. Connect a meter to the channel to be adjusted and set to voltage or
shows the analog output pins and
identifies the associated channels.
2. From the Main Menu, press
to scroll to Service, press
>
to scroll to Analog Out Calibration
,
and press
.
The Analog Output Cal screen displays
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
3. At the Analog Output Cal screen, press
to scroll to the voltage
channel or current channel corresponding to the rear panel terminal
pins where the meter is connected, then press
.
4. With the cursor at Calibrate Zero, press
.
Table 6-5.
Analog Input Channels and Rear Panel Pin Connections
Input Channel
Pin
1
1
2
2
3
3
4
5
5
6
6
7
7
9
8
10
Ground
4, 8, 11, 14