Chapter 10
Performing a Leaks Check (Capillary/Packed/UFM Columns)
Ensuring Tightness
Maintenance and Troubleshooting Manual
187
NOTE
To abort Leak Check, press
STOP
. A relevant message will be displayed.
14. The system is pressurized for one minute, then the pressure value of the
carrier is automatically set to
Off.
The system monitories the pressure for one
minute. During this time, if the pressure
does not drop
more than the
Sensitivity
set value, the message
Leak check passed
is displayed.
R/L.LEAK CHECK
COMPLETED
SUCCESSFULLY
Leak check passed.
If not, the message
Leak detected
is displayed indicating possible leaks in the
system.
15. If leaks are detected, check the accessible, critical connections with an
electronic leak detector to locate the possible leaks.
16. Eliminate the leaks and repeat the leak check procedure.
17. If the leaks cannot be located, contact your customer support organization.
Refer to Appendix B,
, for contact information..