Chapter 12
Maintaining an Electron Capture Detector
Replace the Collecting Electrode
230
Maintenance and Troubleshooting Manual
1. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the ECD, to access
the ECD control table. In the following
example, an ECD installed on the right channel
is considered.
.
RIGHT DETECTOR (ECD)
Base temp
250
Off <
ECD temp
250
Off
Ref current nA
1.0
Freq kHz
(2.15)
Pulse amp V
25
Pulse width us
1.0
Mkup (N2)
40
Off
2. Scroll subsequently to
Base temp
and
ECD
temp
and press
OFF
.
3. Scroll to
Mkup(N2)
and press
OFF
.
4. Disconnect the signal, excitation and heater
cables from the detector.
5. Remove the cover and the detector chimney
unscrewing the two retaining Allen screws. Use
a 1.5 mm Allen wrench.
6. Unscrew the collecting electrode and remove it
using tweezers.