Chapter 12
Wipe Test
Maintaining an Electron Capture Detector
Maintenance and Troubleshooting Manual
223
Wipe Test
Before leaving the factory, each ECD is leak tested for surface radio
contamination by means of the
Wipe Test
method. Each detector is provided with
a
Wipe Test Certificate
reporting the results of the values found and the
procedure followed.
NOTE
The users of the ECD detector in the United States are required to perform a Wipe Test on
their ECD at intervals not exceeding 3 years (36 months), following the reported procedure.
For other Countries, please refer to the appropriate agency for equivalent requirements.