Chapter 3
Replace the Septum
Maintaining a Large Volume Splitless Injector
Maintenance and Troubleshooting Manual
59
5. When the inlet reaches the room temperature,
press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating.
RIGHT CARRIER (S/SL)
Pressure
30.0
30.0
Col. flow
3.00
<
Lin. Veloc.
(60.9)
6. Scroll to
Pressure
and press
OFF
.
7. Unscrew the injector cap.
8. Remove the septum from the septum holder
(use non-metallic tools)