Chapter 9
Maintaining a PTV Injector
Clean or Replace the Liner
156
Maintenance and Troubleshooting Manual
18. Place the septum support into its housing inside
the injector.
19. Reinsert the septum holder with the septum.
20. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
21. Press
RIGHT CARRIER,
scroll to
Pressure
and
press
ON
.
22. Press
OVEN
, scroll to
Temp
and press
ON
.
23. Press
RIGHT INLET,
scroll to
Temp
and press
ON
.
24. Perform a leak check according to
Sequence on page 181.